Transthoracic PacerTest Load: 50 Ohm fixed 12 Selectable steps from 100 to 1000 ohms with EXPMT Load Selector Current: 0.1 to 25 mA Accuracy: +/- 10%, 1 to 10 mA +/- 5%, 10 to 200 mA Pulse Rate: 30 to 800 PPM Accuracy: +/- 1% or 2 ppm Pulse Width: 0.5 to 80 ms Accuracy: +/- 10%, 0.5 to 5 ms +/- 5%, 5 to 25 ms +/- 1%, 25 to 80 ms Sensed and Paced Refractory Period: 25 to 500 ms Accuracy: +/- 10 ms 50/60 Hz Immunity Test: 0 to 10.6 mV in 0.7 mV steps Transvenous PacerTest Load: 500 Ohms +/- 1% Current: 0.1 to 25 mA Accuracy: +/- 10%, 0.1 to 10 mA +/- 5%, 1 to 25 mA Pulse Rate: 30 to 800 PPM Accuracy: +/- 1% or 2 ppm Pulse Width: 0.5 to 80 ms Accuracy: +/- 10%, 0.5 to 5 ms +/- 5%, 5 to 25 ms +/- 1%, 25 to 80 ms Sensed and Paced Refractory Period: 25 to 500 ms Accuracy: +/- 10 ms 50/60 Hz Immunity Test: 0 to 102 mV in 6.8 mV steps Demand SensitivityWaveforms: SQR, TRI, and SSQ Amplitude: 0 to 100 mV Width: 10, 25, 40, 100, and 200 ms Pacer Rate: 30 to 120 ppm A - V Interval0 to 999 ms
|