DELTA 3000 - Defibrillator and External Pacer Analyzer
 |
The DELTA 3000 combines the complete testing functions of a defibrillator tester, a transthoracic pacemaker analyzer, a transvenous pacemaker analyzer and a 12 Lead ECG / Arrhythmia simulator into one compact, lightweight, and easy to use instrument. No other instrument delivers the performance testing power of the DELTA 3000.
The DELTA 3000 tests all defibrillators including AEDs. It measures both monophasic and bi-phasic waveforms. It tests all external pacemakers including transthoracic, transvenous, and AV sequential models. It has 12 built-in user selectable test loads for transthoracic pacing testing. The built-in ECG / Arrhythmia simulator allows performance testing of any ECG equipment.
The DELTA 3000 is operated via easy to follow menus and eight soft keys. Instructions and test measurements are displayed on the large four line LCD display. A print key outputs test measurements to the RS 232 serial port.
The DELTA 3000 is delivered in a hard carrying case with various paddle adapters and interface cables. It provides unsurpassed biomedical testing capabilites in one instrument.
|
|
|
 |
 |
 |
Compact, Lightweight, and Rugged |
 |
Simple, Menu Driven Operation |
 |
Large 4-line LCD |
 |
Simultaneous Display of All Test Measurements |
 |
Dedicated Print Key for Immediate Documentaion |
 |
RS 232 Interface |
 |
Battery or AC Operation |
 |
Two year warranty when annually calibrated by Netech |
|
 |
 |
Measures: Energy, Peak Voltage, Peak Current, and Cardio Sync. Time |
 |
Storage and Playback of output waveforms |
 |
Tests all External Pacemakers - Transthoracic, Transvenous, and AV Pacers |
 |
Measures: Amplitude, Rate, Pulse Width, Pulse Energy, Sensed and Paced Refactory Periods |
 |
50 and 60 Hz Immunity Tests |
 |
Transthoracic Test Loads: Fixed 50 Ohm or 12 User Selectable Loads to 1000 ohms |
 |
A / V Pacer Test Load: 500 Ohm |
 |
Tests all ECG Equipment. |
 |
12 lead ECG, User Selectabel Rates and Amplitudes |
 |
Performance Waveforms: Sine, Square, Triangle amp; Pulse |
 |
12 Arrhythmias |
|

|
|
Defibrillator AnalyzerEnergy Measurement:Load: 50 Ohm +/- 1 % Range (High): 900 J Resolution: 1 Joule Range (Low): 0 - 100 Resolution: 0.1 Joules Accuracy: +/- 2% of the reading + 1 J for Hi Range or + 0.5 J for Low Range Maximum Voltage: 5200 V Maximum Current: 100 Amp (H) Sync. Time: 0 - 250 ms Waveform Output:Real Time: Output 1000 : 1 Stored. Discharge waveform expanded 200:1 and is repeated 8 sec. It can be viewed through the paddles and through the ECG High Level Output Pacemaker AnalyzerTranscutaneous:Test Load: 50 Ohm (Fixed) and variable 100 to 1000 Ohms user selectable Current: 1 to 200 mA Accuracy: +/- 5% + 0.5 mA Pulse Width: 0.5 to 80 ms Accuracy: +/- 1% + 0.1 ms Refractory: 20 - 500 ms Accuracy: +/- 2% or + 10 ms Atrial and Ventricular:Test Load: 500 Ohms Current: 1 to 25 mA Accuracy: +/- 5% + 0.5 mA Pulse Rate: 30 to 800 ppm. Accuracy: +/- 1% + 0.1 ms Refactory: +/- 20 - 500 ms Accuracy: +/- 2% or 10 ms Immunity Test: 50 / 60 Hz ECG ArrhythmiaSimulatorECG: Normal Sinus Rhythm.Rate: 30, 60, 90, 120, 240 Performance Waveforms:Sine, Square, Triangle and Pulse Rate: 0.5, 1, 2, 10, 15, 20, 25, 40 Hz Arrhythmias:VFBC, VFBF, VTAC, AFIB, PVC, PVC1, PVC2, BGY, RBB, AFIB, ATRFT, NSR Waveform Output: 12 Lead ECG and Paddle and High Level Output Jacks. 
|
Power Requirements:Two 9V Alkaline Batteries Or AC Adapter  Dimension:10 x 11 x 2.25 inches (25.4 x 27.9 x 5.7 cm) Weight: 4 lbs (1.8 kg) |
|
|
 |
|

 |