ECG Arrhythmia Simulator

Product Code : MiniSim 1000 ECG, Arrhythmia Waveform Simulator

 MiniSim-1000 Arrhythmia Simulator is a powerful yet basic patient simulator in a compact case. It is designed for testing the performance of all ECG monitoring instrumentation quickly and easily. 300his comprehensive microprocessor-based instrument is menu driven using the tactile feel keypad and easy-to-follow LCD display.

The MiniSim-12 provides full 12 lead ECG simulation with 14 user selectable rates ranging from 30 to 350 BPM, and 14 user selectable amplitudes from 0.15 to 5 mV. The performance waveform includes Sine, Square and Triangle . 14 user-selectable waveform rates from 0.1 to 100 Hz and amplitudes from 0.15 millivolt to 5 millivolts.

The instrument is also designed to simulate 45 different arrhythmias including atrial, ventricular, tall T wave, ST elevation, ST depression, myocardial infarction, blocks, pulse and pacer waveforms.

 MiniSim 1000 Arrhythmia Simulator is software upgradable to MiniSim 1000 full function patient simulator including Blood pressure, respiration, and temperature (YSI 400/YSI 700 compatible) 

 

Show More... Add to cart

    • Best Value
    • Easy to Use
    • Compact Size
    • Menu Driven operation
    • Large LCD Display
    • Tactile Keypad
    • Battery or AC Operation

    ECG: 12 leads with independent outputs referenced to RL
    NORMAL SINUS RHYTHM (NSR)
    ECG Rates: 30, 60, 70, 80, 90, 100, 120, 150, 180, 210, 240, 270, 300, and 350 BPM. Accuracy 0.5%.
    Amplitudes: 0.15, 0.3, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 mV on Lead II. Lead 1 is 0.6 X Lead II, Lead III is 0.4 X Lead II, and Lead V is the same as Lead II.
    High Level: 500 X low-level output on Lead II.
    Accuracy: 2% (1-5mV).

    PERFORMANCE WAVEFORMS (PERF)

    Sine, Square, Triangle, Pulse
    Frequencies: 0.1 to 0.9 in 0.1 Hz steps.1.0 to 9.0 in 1.0 Hz steps.10 to 100 in 10 Hz steps.
    Accuracy: 1%
    Amplitudes: 0.1, 0.2, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 mV on Lead II.
    Pulse: 20 ms pulse of 1mV amplitude repeated at 4-second intervals.

    R WAVE DETECTION (RWD)

    Widths: 10, 40, 50, 60, 70, 80, 90, 100, 110, and 120 ms.
    Amplitudes: 0.15, 0.3, 0.5, 1.0, 2.0, 3.0, 4.0, 5.0 mV on Lead II
    PACER (PCR)
    AP: Atrial Pacer set at 70 BPM
    ASP: Asynchronous Pacer- Ineffective Pacing
    QRS Amplitudes: 1 mV
    PCR: Pacer pulses alone
    Pacer Widths: 0.1, 0.2, 0.5, 1.0, 1.5, and 2.0 ms
    Amplitudes: 0.15, 0.3, 0.5, 1.0, 2.0, 3.0, 4.0, and 5.0 mV
    VP: Ventricular Pacer set at 70 BPM
    AVP: Atrial Ventricular Pacer set at 70 BPM
    QRS Amplitude: 1 mV.Pacer Amplitude: -2mV

     

Download Brochure Download Manual
Part Number Description Price
300-0.1 _MiniSim-12 lead ECG and Arrhythmia Simulator Add to cart
300-0.1-UPGRADE MiniSim 1000 Arrhythmia simulator optional upgrade to Full function Patient Simulator 300-1.3 Add to cart